At present, most of the orthopaedic implants used in articular reconstruction are fixed to host bone using acrylic bone-cement. Bone-cement polymerization leads to an exothermic reaction with heat release and consequent temperature rise. The increase of temperature in the bone beyond the tolerated limits can develop osteocyte thermal necrosis and ultimately lead to bone resorption at the cement-bone interface, with subsequent loosening of the implant. Another issue that plays an important role in implant loosening is debonding of the cement from the implant initiated by crack formation at the interfacial voids. It is well established that low porosity enables better fatigue cement properties. Moderate preheating of the implant is expected to reverse the direction of polymerization, and has the ability to reduce interfacial void formation and improve interfacial shear strength. To increase the implant temperature at the initial cementing phase in order to reduce interfacial void formation, and subsequently, cool the implant in the latter cement polymerization phase to prevent the possibility of bone thermal necrosis, a new automated electronic device was designed to be use in cemented joint replacements. The developed device was specifically designed for the knee arthroplasty, namely for tibial-tray cementing. The device controls the heat flux direction between the tibial-tray and the atmosphere through the “Peltier effect,” using Peltier tablets. The device is placed on the tibial-tray during the cementing phase and starts to heat it in a first phase, promoting the polymerization that initiates at the warmer cement-implant interface. In a second phase, the heat flux in the Peltier tablets is inverted to extract the heat generated during cement polymerization. The device efficiency was evaluated by cementing several tibial-trays in bovine fresh bone and measuring the tray and cement temperatures. The temperature results in the implant and in the cement showed that the device increases and maintains the implant temperature above room temperature at the initial cementing phase, while in the subsequent phase it cools the tibial-tray and cement. Significant differences were found for peak cement temperatures between the tests performed with and without the device. The device showed its capacity to promote the beginning of cement polymerization at the implant interface contributing towards improving interfacial shear strength and in reducing the peak cement temperature in the subsequent polymerization process, thus contributing to the prevention of the bone thermal necrosis effect.
Skip Nav Destination
e-mail: completo@ua.pt
Article navigation
March 2012
Technical Briefs
A Device to Control Implant and Bone-Cement Temperatures in Cemented Arthroplasty
A. Completo,
A. Completo
Department of Mechanical Engineering,
e-mail: completo@ua.pt
University of Aveiro
, 3810-193 Aveiro, Portugal
Search for other works by this author on:
M. Coutinho,
M. Coutinho
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
Search for other works by this author on:
M. Schiller,
M. Schiller
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
Search for other works by this author on:
A. Ramos,
A. Ramos
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
Search for other works by this author on:
C. Relvas,
C. Relvas
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
Search for other works by this author on:
J. A. Simões
J. A. Simões
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
Search for other works by this author on:
A. Completo
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
e-mail: completo@ua.pt
M. Coutinho
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
M. Schiller
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
A. Ramos
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
C. Relvas
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
J. A. Simões
Department of Mechanical Engineering,
University of Aveiro
, 3810-193 Aveiro, Portugal
J. Med. Devices. Mar 2012, 6(1): 014503 (4 pages)
Published Online: March 13, 2012
Article history
Received:
April 27, 2011
Revised:
November 11, 2011
Published:
March 12, 2012
Online:
March 13, 2012
Citation
Completo, A., Coutinho, M., Schiller, M., Ramos, A., Relvas, C., and Simões, J. A. (March 13, 2012). "A Device to Control Implant and Bone-Cement Temperatures in Cemented Arthroplasty." ASME. J. Med. Devices. March 2012; 6(1): 014503. https://doi.org/10.1115/1.4005783
Download citation file:
36
Views
Get Email Alerts
Cited By
Related Articles
A Device to Control Implant and Bone-Cement Temperatures in Cemented Arthroplasty
J. Med. Devices (June,2011)
Finite Element Thermal Analysis of Bone Cement for Joint Replacements
J Biomech Eng (June,2003)
An Analysis of Crack Propagation Paths at Implant/ Bone-Cement Interfaces
J Biomech Eng (November,1996)
The Compatibility of Thin Films and Nanostructures in Thermoelectric Cooling Systems
J. Heat Transfer (July,2007)
Related Proceedings Papers
Related Chapters
Thermoelectric Coolers Are Hot
Hot Air Rises and Heat Sinks: Everything You Know about Cooling Electronics Is Wrong
Thermoelectric Coolers
Thermal Management of Microelectronic Equipment
Comparative Study of Carbon Polymer Composite and Titanium Femoral Stems in Dogs Using Computed Tomography
Composite Materials for Implant Applications in the Human Body: Characterization and Testing